€ 48,90
Friction and lubrication behaviour of 50mm diameter Birmingham Hip Resurfacing implants with diametral clearances 80 to 300µm, using blood (clotted and whole blood), a combination of bovine serum with hyaluronic acid and carboxymethyl cellulose (CMC) adjusted to a range of viscosities (0.001-0.2 Pas), and bovine serum with CMC adjusted to a similar range of viscosities, have been investigated. The result of this investigation has suggested strongly that the optimum clearance for the 50 mm diameter MOM BHR implants to be ≥150µm and <235µm, so as to avoid high frictions (i.e. avoid friction factors >0.2) and be able to accommodate a mixed lubrication mode and hence lower the risk of micro- or even macro-motion specially immediately after hip implantation. These suggested optimum clearances will also allow for low friction (i.e. friction factors of <0.2-0.07) and reasonable lubrication (dominantly mixed regime) for the likely cup deflection occurring as a result of press-fit fixation.